Condition Comments:
Cosmetic Condition:
The reader is in generally good cosmetic condition despite the "Cuvette Read" button being torn (button still depresses and the command is recognized). The instrument initially failed the self test at startup, presenting the error message "#313 Call Support". Instrument now powers on successfully. Testing the instrument with a Molecular Devices ABS1 Absorbance Validation Test Plate provided the pictured results. Instrument is able to follow temperature control accurately.